Escudo de la República de Colombia
Sistema Nacional de Biliotecas - Repositorio Institucional Universidad Nacional de Colombia Biblioteca Digital - Repositorio Institucional UN Sistema Nacional de Bibliotecas UN

Estudio de validez diagnóstico: consistencia del dinamómetro de mano digital Camry en una población de adultos sanos en Bogotá

Diaz Muñoz, Gustavo Alfonso (2016) Estudio de validez diagnóstico: consistencia del dinamómetro de mano digital Camry en una población de adultos sanos en Bogotá. Maestría thesis, Universidad Nacional de Colombia - Sede Bogotá.

Texto completo

[img]
Vista previa
PDF - Versión Aceptada
Available under License Creative Commons Attribution Non-commercial No Derivatives.

2MB

Resumen

Objetivo: Evaluar la concordancia-consistencia del dinamómetro Camry comparado con el dinamómetro Jamar en una población de adultos de 18 o más años de edad en Bogotá. Metodología: Estudio observacional tipo concordancia-consistencia en adultos de ≥18 años, residentes en Bogotá y libres de enfermedades que alteren la medición de fuerza; se compararon el dinamómetro Camry Vs. el Jamar. Se utilizó el coeficiente de correlación concordancia (IC95% > 0.9) y el método gráfico de Bland-Altman. Resultados: Participaron 133 personas (50% mujeres), entre 18-88 años y edad promedio 47 años (+/- 20,7). La fuerza promedio en la población con el dinamómetro Jamar fue 32,15kg (+/-9,96 mano derecha) y 30,43kg (+/-9,81 mano izquierda); con el dinamómetro Camry fue 29,95 kg (+/-9,18) y 28,2 kg (+/- 9,02) para la mano derecha e izquierda respectivamente. Se obtuvieron CCC con significación estadística a nivel poblacional (CCC 0,933 IC95% 0,911-0,955 en ambas manos) y en el grupo de 40-59 años (CCC 0,96 IC95% 0,934-0,985 mano derecha; CCC 0,964 IC95% 0,934-0,981 mano izquierda). Los gráficos de Bland-Altman mostraron que la mayoría de los datos están entre los límites de concordancia y la diferencia media resulto en todos los casos con valores negativos; los límites de concordancia en mujeres fue -5,95-2,93 kg y en hombres -8,96-3,4 kg. Conclusiones: El equipo Jamar es intercambiable por el equipo Camry en el grupo de edad de 40-59 años; el equipo Camry podría ser un proxy del Jamar en la práctica clínica, lo cual debe ser evaluado en futuras investigaciones., Abstract: Objective: Evaluate the concordance-consistency of the Camry dynamometer Vs. Jamar dynamometer in adults of 18 years or older in Bogotá. Methodology: Observational concordance-consistency study in adults aged ≥18 years, living in Bogota and free of diseases that alter strength measurement; The Camry Vs. the Jamar dynamometer was compared. Statistical analysis was with Correlation Coefficient Concordance (95% CI> 0.9) and the Bland-Altman graphical method. Results: Participated 133 people (50% women), between 18-88 years and average age 47 years (+/- 20.7). The average handgrip strength in the population with the Jamar dynamometer was 32.15kg (+/- 9.96 right hand) and 30.43kg (+/- 9.81 left hand); With the Camry dynamometer was 29.95 kg (+/- 9.18) and 28.2 kg (+/- 9.02) for the right and left hand respectively. The CCC were statistically significant at the population level (CCC 0.933 CI95% 0.911-0.955 in both hands) and in the group of 40-59 years (CCC 0.96 CI95% 0.934-0.985 right hand; CCC 0.964 IC95% 0.934-0.981 hand left). The Bland-Altman graphs showed that most of the data are within the bounds of agreement and the mean difference resulted in all cases with negative values; The limits of agreement were -5.95 a 2.93 kg and -8.96 a 3.4 kg for women and men respectively. Conclusions: The Jamar dynamometer is interchangeable with the Camry dynamometer in the age group of 40-59 years; The Camry dynamometer could be a proxy for Jamar in clinical practice, which should be evaluated in future research.

Tipo de documento:Tesis/trabajos de grado - Thesis (Maestría)
Colaborador / Asesor:Rodriguez Godoy, Mauricio
Información adicional:Magíster en Epidemiología Clínica
Palabras clave:Dinamómetro de Fuerza Muscular, Fuerza de la Mano, Reproducibilidad de Resultados, Adulto, Mediana Edad, Estudios Transversales, Anciano, Muscle Strength Dynamometer, Strength Hand, Reproducibility of Results, Adult, Middle Aged, Aged, Cross-Sectional Studies
Temática:5 Ciencias naturales y matemáticas / Science > 51 Matemáticas / Mathematics
6 Tecnología (ciencias aplicadas) / Technology > 61 Ciencias médicas; Medicina / Medicine & health
Unidad administrativa:Sede Bogotá > Facultad de Medicina > Instituto de Investigaciones Clínicas
Código ID:54338
Enviado por : Sr gustavo alfonso díaz muñoz
Enviado el día :24 Noviembre 2016 23:10
Ultima modificación:24 Noviembre 2016 23:10
Ultima modificación:24 Noviembre 2016 23:10
Exportar:Clic aquí
Estadísticas:Clic aquí
Compartir:

Solamente administradores del repositorio: página de control del ítem

Vicerrectoría de Investigación: Número uno en investigación
Indexado por:
Indexado por Scholar Google WorldCat DRIVER Registry of Open Access Repositories OpenDOAR Metabiblioteca BDCOL OAIster Red de repositorios latinoamericanos DSpace BASE Open archives La referencia Colombiae Open Access Theses and Dissertations Tesis latinoamericanas CLACSO
Este sitio web se ve mejor en Firefox